Progress in Coherence Imaging Techniques on DIII-D

POSTER

Abstract

Coherence Imaging Spectroscopy CIS systems are used on DIII-D for measurement of 2D distributions of impurity ion flow velocities, Imaging Motional Start Effect-(IMSE), and a newly developed impurity ion temperature diagnostic [1]. High-quality calibrated flow images previously required careful temperature control; rapid inter-shot system calibration with a laser and wave meter has reduced this requirement. The tunable laser has facilitated CIS characterization and calibration, and results from interferometers with three different optical delays (1000-flow, 1300, 2900-ion temperature waves) will be presented. More than one emission line in the CIS bandpass requires extra analysis; an example with two emission lines is shown. Calibrated tomographically inverted CIII flow images of the DIII-D divertor plasma will be presented. In the case of IMSE, the fidelity of the measurement depends on a switchable polarization retarder being a true half-wave. We have worked to improve the liquid crystal variable retarder and initial results will be shown. [1] C.M. Samuell, et al., these proceedings

Authors

  • S.L. Allen

    LLNL, Lawrence Livermore National Lab

  • C.M. Samuell

    LLNL, Lawrence Livermore National Lab

  • B S Victor

    LLNL, Lawrence Livermore National Lab, Lawrence Livermore Natl Lab