Updates to the FAIRRAY Detector at the UMass Lowell Research Reactor
ORAL
Abstract
The FAIRRAY Detector for high-resolution measurements of gamma-rays from thermal neutron-capture reactions at the University of Massachusetts Lowell Research Reactor (UMLRR) has been updated with improved detectors and down-chain electronics. The addition of transistor-reset preamplifiers (TRP) to the high-purity germanium (HPGe) detectors improve performance in high count-rate environments, and a custom filter and amplifier circuit was developed to interface with the existing DAQ hardware. We present these modifications, additional solutions implemented for dead-time correction, and future plans for the detector array.
–
Presenters
-
Daniel E Fernandez
University of Massachusetts Lowell
Authors
-
Daniel E Fernandez
University of Massachusetts Lowell
-
Marian Jandel
University of Massachusetts Lowell
-
Stanislav Valenta